SBX-75
and SBX-75A Ionization Unit In Field Test
QUICK
IONIZATION UNIT FIELD TEST
- Disconnect
and remove all power to the unit.
- Put the test/operate
jumper in the test position (lower two horizontal pins)on the
main circuit board.
- Check the
fuse with a multimeter for continuity.
- Reconnect
and turn power to the unit on.
- Turn the
power switch on.
- The Power
Monitor light will alternate between green and orange every 9-10 seconds.
On older units the red Power Monitor light will alternate between
on and off.
- Disconnect
the wires from the electrodes and short them together.
- Rotate the
ION OUTPUT knob to 30%.
- The ION OUTPUT
light should illuminate. Short the leads for
less than 15 seconds!
- Remove the
short between the electrode wires - the CHECK ELECTRODES light should illuminate.
- Turn
the power off.
- Reconnect
the wires to the electrodes. Rotate the ION OUTPUT knob to 0%.
IONIZATION
OUTPUT TEST WITH MULTIMETER
It
is mandatory to follow the multimeter manufacturer's directions and safety instructions.
Note: For ease
of testing, we suggest the multimeter should be a digital auto-ranging device
with auto polarity.
- Turn
all power to the unit off.
- Turn Ion
Output knob to the "0" or "OFF" position.
- Disconnect
one lead from unit to the electrodes, at the electrodes. Connect the meter
as shown in below:
- Set switches
on meter to measure 10 or 20 amps DC current
- Verify leads
to the meter are hooked up properly, the 10 or 20 amp scale usually requires
the leads to be in special positions on the meter.
- Turn on power
to the unit. Rotate the timer until the pump turns on.
- Meter should
read less than 0.02 Amps . (20 ma)
- Turn Ion
Output knob to "30".
- Meter should
read between 0.1 and 0.2 amps. (100 and 200 ma)
- Turn Ion
Output knob to "100" (full cw).
- Meter should
read between 0.40 and 0.60 amps (400 and 600 ma). Note: on old units the reading
may be up to 0.80 . (800 ma)
- Wait about
3 1/2 minutes and polarity should change. (+ and -)
- If the unit
is in the test mode; the polarity will change more rapidly.
- Turn power
off to the unit. Turn power off on meter. Disconnect test leads.
- Return electrode
lead to the Ionization Chamber.
- Return ionization
output knob to "0" or previous setting.
If the unit
does not pass these tests, note the step and failure. Contact
the factory for further insturctions.
This completes the Field
Test Instructions for the SBX-75 and SBX-75A units.