WARNING:
USE EXTREME CAUTION IN THE USE OF THE MULTIMETER. IMPROPER USE OF THIS DEVICE CAN RESULT IN INJURY OR DEATH. FOLLOW ALL SAFEGUARDS SUGGESTED IN THE OWNERS MANUAL FOR THE MULTIMETER IN ADDITION TO NORMAL SAFETY PRECAUTIONS IN DEALING WITH ELECTRICAL CIRCUITS. DO NOT USE THIS DEVICE IF YOU ARE UNFAMILIAR WITH ELECTRICAL CIRCUITS AND TESTING PROCEDURES.
FOLLOW THE METER MANUFACTURERS INSTRUCTIONS CAREFULLY AND COMPLETELY. OBSERVE AND FOLLOW ALL CAUTION WARNINGS!
NOTE: For ease of testing, we suggest the multimeter should be a digital auto-ranging device with auto polarity.
Go
straight to the troubleshooting section.
[ ] Before starting the test, check the following:
[ ] Turn all power to the unit OFF. On Model 5001 the Auto/Manual switch must be in the "Auto" position.
[ ] Turn Purifier Output knob to the "0" or the "OFF" position.
[ ] Disconnect one lead from unit to the electrodes, at the electrodes. Connect the multimeter and the calibrating resistor (if applicable) as shown in Figure 1. The resistor and test lead can be obtained from Wailani at no cost less shipping. (Order calibrating resistor.)
Figure 1
[ ]
Set switches on the meter to measure 10 or 20 amps DC current.
Verify leads to the meter are hooked up properly, the 10 or 20 amp scale usually
requires the leads to be in special positions.
[ ]
For Model 1500 and 2001 - Turn the pump on and the unit should turn on.
For Model 5001 - Turn on the power to the unit. Place the Auto/Manual switch
in the Manual position. The pump should turn on. Verify "flow" (actually
a no flow indicator) light is off.
Meter should read less than 0.02 Amps . (20 ma)
[ ]
On the Model 1500 set the output to 30%.
On the 2001 and 5001, turn the Output knob so that the 50% light is illuminated
on the front panel.
For a Model
1500 unit the reading should be between .070 and .125 (70 and 125 ma)
On the 2001 and 5001 the meter should read between 0.2 and 0.3 amps. (200 and
300 ma)
On the 2051 or High Output purifier the meter should read between 0.3 and 0.4
amps.(300 and 400ma)
[ ] Turn Output knob to "100" (full cw).
The Model
1500 should read between 0.30 and 0.40 (300ma and 400ma).
For the 2001 and 5001 units, the meter should read between 0.450 and 0.525 amps.
(450 to 525 ma)
For the 2051 or high Output purifier the meter should read between 0.700 and
0.800 amps. (700 to 800 ma)
[ ]
Wait about 3 1/2 minutes and polarity on the meter should change. (+ and
-). The Power indicator will change color about every 3 1/2 minutes.
( If the unit is in the test mode; the polarity will change more
rapidly.)
[ ] Turn the power off to the unit. Disconnect test leads and resistor.
[ ] Return electrode lead to the Purifier chamber.
[ ] Return output knob to "0" or previous setting.
Connect the system and the calibrating resistor as shown below in Figure 2. Retest the system per the instructions above.

Figure 2
If the tests now pass there is a problem with the electrode assembly.
If reading is always less than 0.01, then no ion current is flowing:
If output is lower than specified:
If there is a reading for 3 1/2 minutes, then it goes to zero or does not change polarity:
If the above measures do not repair the unit; then contact Wailani Water Products.
| Previous | Home Page | |
| ©2003-2008 Wailele Pools Inc. All Rights Reserved |